Description
Standard-Incubators with natural convection Model BD 260 Binder Germany
-
Benefits
- Excellent temporal and spatial temperature accuracy
- High process reliability
- 100°C disinfection routine
Product description
The BD Avantgarde.Line series is specially designed for gentle incubation of organisms, as well as conditioning of heat-sensitive media. The new design of the units and the homogeneous temperature distribution ensure the very best results.
Important characteristics
- Temperature range: +5°C above ambient temperature to +100°C
- High temperature accuracy thanks to APT.line™ technology
- Natural convection
- Controller with LCD display
- Electromechanical control of the exhaust air flap
- Inner door made of tempered safety glass
- 2 chrome-plated racks
- Class 3.1 integrated independent temperature safety device (DIN 12880) with visual alarm
- Ergonomic handle design
- USB port for recording data
- Technical specifications
Data
| Designation |
BD260-230V |
BD260UL-120V |
| Article Number |
9010-0329 |
9010-0330 |
| Option model |
Standard |
Standard |
Performance Data Transfer
| Designation |
BD260-230V |
BD260UL-120V |
| Temperature range |
+5 °C above ambient temperature to 100 °C |
+5 °C above ambient temperature to 100 °C |
| Temperature uniformity at 37°C |
0.3 ± K |
0.3 ± K |
| Temperature uniformity at 100°C |
1.0 ± K |
1.0 ± K |
| Temperature fluctuation at 37°C |
0.2 ± K |
0.2 ± K |
| Temperature fluctuation at 100°C |
0.5 ± K |
0.5 ± K |
| Heating up time to 100°C |
75 min |
75 min |
| Heating up time to 37°C |
60 min |
60 min |
| Recovery time after door was opened for 30 s at 37°C |
16 min |
16 min |
Electrical Data
| Designation |
BD260-230V |
BD260UL-120V |
| Rated Voltage |
230 V |
120 V |
| Power frequency |
50/60 Hz |
60 Hz |
| Nominal power |
0,85 kW |
0,95 kW |
| Unit fuse |
8,0 A |
12,5 A |
| Phase (Nominal voltage) |
1~ |
1~ |
Dimensions and Weights
| Designation |
BD260-230V |
BD260UL-120V |
| Interior volume |
253L |
253L |
| Net weight of the unit (empty) |
85kg |
85kg |
| Load per rack |
40kg |
40kg |
| Permitted load |
270kg |
270kg |
| Wall clearance back |
160mm |
160mm |
| Wall clearance sidewise |
100mm |
100mm |
Housing Dimensions not incl. fittings and connections
| Designation |
BD260-230V |
BD260UL-120V |
| Width net |
810 mm |
810 mm |
| Height net |
940 mm |
940 mm |
| Depth net |
760 mm |
760 mm |
Internal Dimensions
| Designation |
ED260-230V |
ED260UL-240V |
| Interior width |
610 mm |
610 mm |
| Interior height |
760 mm |
760 mm |
| Interior depth |
545 mm |
545 mm |
| Inner doors |
1 |
1 |
| Unit doors |
1 |
1 |
Environment-specific data
| Designation |
BD260-230V |
BD260UL-120V |
| Energy consumption at 100°C |
225 Wh/h |
225 Wh/h |
| Energy consumption at 37°C |
40 Wh/h |
40 Wh/h |
Fixtures
| Designation |
BD260-230V |
BD260UL-120V |
| Number of shelves (std./max.) |
2/8 |
2/8 |