Drying and heating chambers with gravity convection Model ED 56Drying and heating chambers with gravity convection Model ED 56 Binder Germany Binder Germany
-
Benefits
- Best energy efficiency on the market
- Excellent temporal and spatial temperature accuracy
- USB connection for recording data
- Temperature range up to +300 °C
Product description
The strengths of a BINDER ED series drying chamber include routine drying and sterilization tasks up to +300 °C. Thanks to the natural convection, all thermal processes which use this drying chamber are highly efficient. The ED series ensures fast, uniform drying.
Important characteristics
- Temperature range: ambient temperature plus 5 °C to 300 °C
-
- High temperature accuracy thanks to APT.line™ technology
- Gravity convection
- Controller with LCD display
- Electromechanical control of the exhaust air flap
- 2 chrome-plated racks
- Stackable
- Class 2 independent temperature safety device (DIN 12880) with visual temperature alarm
- Ergonomic handle design
- USB port for recording data
Technical specifications
Data
Designation |
ED056-230V |
ED056UL-120V |
Article Number |
9010-0333 |
9010-0334 |
Option model |
Standard |
Standard |
Performance Data Transfer
Designation |
ED056-230V |
ED056UL-120V |
Temperature range |
+5 °C above ambient temperature to 300 °C |
+5 °C above ambient temperature to 300 °C |
Temperature uniformity at 150°C |
2.3 ± K |
2.3 ± K |
Temperature fluctuation at 150°C |
0.4 ± K |
0.4 ± K |
Heating up time to 150°C |
50 min |
50 min |
Recovery time after door was opened for 30 s at 150°C |
19 min |
19 min |
Air Change Data
Designation |
ED056-230V |
ED056UL-120V |
Air change rate at 100°C |
8 x/h |
8 x/h |
Electrical Data
Designation |
ED056-230V |
ED056UL-120V |
Rated Voltage |
230 V |
120 V |
Power frequency |
50/60 Hz |
60 Hz |
Nominal power |
1,05 kW |
1,15 kW |
Unit fuse |
6,3 A |
12,5 A |
Phase (Nominal voltage) |
1~ |
1~ |
Dimensions and Weights

Housing Dimensions not incl. fittings and connections

Internal Dimensions

Environment-specific data
Designation |
ED056-230V |
ED056UL-120V |
Energy consumption at 100°C |
105 Wh/h |
105 Wh/h |
Energy consumption at 150°C |
180 Wh/h |
180 Wh/h |
Fixtures
Designation |
ED056-230V |
ED056UL-120V |
Number of shelves (std./max.) |
2/4 |
2/4 |