Ultra-low temperature freezers with climate-neutral refrigerants Model UF V 350 Binder Germany
Call: 01715 130348
Ultra-low temperature freezers with climate-neutral refrigerants
Model: UF V 350
Brand: Binder
Country of Origin: Germany
Description
Ultra-low temperature freezers with climate-neutral refrigerants Model UF V 350 Binder Germany
-
Benefits
- Lowest energy consumption in its class
- Environmentally-friendly refrigerants R-290 and R-170
- Thermal insulation with unique long-life VIPs (vacuum insulation panels)
- Rust-proof interior made completely of stainless steel
Product description
The BINDER ultra low temperature freezer ensures the safe storage of samples at -90 °C. It combines outstanding environmental friendliness with low energy consumption, convenient operation and an individual safety concept.
Important characteristics
- Temperature range:-90 °C to -40 °C
- Lowest energy consumption in its class
- Powerful cascade compressor cooling unit
- Environmentally friendly refrigerants R-290 and R-170
- Thermal insulation with unique long-life VIPs (vacuum insulation panels)
- Innovative door gasket concept reduces buildup of ice
- Ergonomic door handle
- Rust-proof interior made completely of stainless steel
- Detachable inner doors made of stainless steel
- 3 stainless-steel shelves providing flexible placement
- Zero-voltage alarm contact for in-house alarm system
- Internal data logger, measured values can be read out in open format via USB
- Ethernet interface
- 3 access ports, Ø 28 mm, rear
Technical specifications
Performance Data Temperature
Designation |
UFV350-230V |
UFV350UL-120V |
UFV350UL-240V |
Temperature range |
-90…-40 °C |
-90…-40 °C |
-90…-40 °C |
Temperature uniformity at -80°C |
3.5 ± K |
3.5 ± K |
3.5 ± K |
Temperature fluctuation at -80°C |
4.0 ± K |
4.0 ± K |
4.0 ± K |
Cooling down time from 22°C to -80°C |
210 min |
210 min |
210 min |
Electrical Data
Designation |
UFV350-230V |
UFV350UL-120V |
UFV350UL-240V |
Rated Voltage |
230 V |
115 V |
208…230 V |
Power frequency |
50 Hz |
60 Hz |
60 Hz |
Nominal power |
1,6 kW |
1,4 kW |
1,8 kW |
Unit fuse |
10,0 A |
15,0 A |
10,0 A |
Phase (Nominal voltage) |
1~ |
1~ |
2~ |
Dimensions and Weights

Housing dimensions not incl. fittings and connections

Internal Dimensions

Environment-Specific Data
Designation |
UFV350-230V |
UFV350UL-120V |
UFV350UL-240V |
Sound-pressure level |
47 dB(A) |
47 dB(A) |
47 dB(A) |
Energieverbrauch bei Sollwert -80 °C und Tᵤ = 20 °C |
7.5 kWh/d |
7.5 kWh/d |
7.5 kWh/d |
Designation |
UFV350-230V |
UFV350UL-120V |
UFV350UL-240V |
Number of shelves (std./max.) |
3/11 |
3/11 |
3/11 |
Load Variants