The universally applicable lab oven U is Memmert’s classic appliance for temperature control in science, research and material tests in industry. The technologically perfected masterpiece made of high-quality, hygienic, easy-to-clean stainless steel leaves nothing to be desired in terms of ventilation and control technology, overtemperature protection and precisely tuned heating technology.
On this page, you can find all the essential technical data on the universal Memmert lab oven. Our customer relations team will be pleased to help if you want further information.
Temperature
Working temperature range |
at least 5 (UN/UNplus/UNm/UNmplus) or 10 (UF/UFplus/UFm/UFmplus) above ambient temperature to +300 °C |
Setting accuracy temperature |
up to 99.9 °C: 0.1 / from 100 °C: 0.5 |
Setting temperature range |
+20 to +300°C |
Temperature sensor |
1 Pt100 sensor DIN class A in 4-wire-circuit |
Control technology
ControlCOCKPIT |
SingleDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with high-definition TFT-colour display |
Timer |
Digital backwards counter with target time setting, adjustable from 1 minute to 99 days |
Function SetpointWAIT |
the process time does not start until the set temperature is reached |
Calibration |
three freely selectable temperature values |
adjustable parameters |
temperature (Celsius or Fahrenheit), air flap position, programme time, time zones, summertime/wintertime |
Ventilation
Convection |
natural convection |
Fresh air |
Admixture of pre-heated fresh air by electronically adjustable air flap |
Vent |
vent connection with restrictor flap |
Communication
Documentation |
programme stored in case of power failure |
Programming |
AtmoCONTROL software for reading out, managing and organising the data logger via Ethernet interface (temporary trial version can be downloaded). USB stick with AtmoCONTROL software available as accessory (on demand). |
Safety
Temperature control |
adjustable electronic overtemperature monitor and mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature |
Autodiagnostic system |
for fault analysis |
Standard equipment
Works calibration certificate |
Calibration at +160°C |
Door |
fully insulated stainless steel door with 2-point locking (compression door lock) |
Internals |
2 stainless steel grid(s), electropolished |